Thanks for the insight, rgbenoit. I have a question, though: If this is a home monitoring device, as you suggest, would it still have value? Can home users understand an ECG?
This is a very good start with respect to miniaturization... but there are several things readers need to keep in mind... a lot of things actually, for acceptance in the USA in clinical environments as an approved medical device for patient care. There are many short comings with the current device, and the likelihood of acceptance in the USA (both FDA and product market acceptance) would be extremely questionable. However, there would be a good application for this technology in the home monitoring, military and in athletics, where basic EKG data would be nice to have, just not as diagnostic (this already exists for such purposes using ANT+ in common basic physiological monitoring equipment.. watches and HR/breathing), but is very cheap and easy to implement... but WILL NEVER BE USING IN CLINICAL SETTINGS IN THE USA.
First... a little bit about ECGs (EKGs in the US) and sensors. One of the primary disadvantages of only using one electrode to detect the minute electrical signals of the heart is that those signals are highly directional. In fact, there are 12 general directions monitored in what are called 12 Lead EKGs (the gold standard... however, most real-time monitoring is of only one "lead", which is associated with one of the 12 general directions. Unfortunately, using just a single leads is not going to give any "sense of direction" and would be just a simple summary of the electrical activity all the leads. This impedes the ability of the intelligent analysis systems to determine arrhythmia changes (changes to the baseline EKG) beyond those specific to general time-related analysis and the application of chaos theory to detect sudden cardiac death for example, which is well documented in the medical literature. The primary 6 leads, are called limb leads, and are used for basic cardiac monitoring and provide longitudinal electrical "axis" data and better wave form analysis for different "parts" of an EKG. These are the classic 3 or 4 wire "patches" seen on the 4 corners of the chest. 3 wire systems (3 patches) can only record a single limb lead. All 4 patches or limb attachments (back in the old days the "limb" leads were actually your arms and legs) must be present to record all 6 limb leads... which provide about 85% of the diagnostic data needed to do a full cardiac electrical analysis. Note that automatic defibrillators use one limb lead for analysis (across 2 wires). The remaining 6 leads of a 12 lead EKG are called cardiac, or chest leads. They record data from immediately around the heart, to provide a more of a 3D image, as the heart doesn't lie flat in the chest and tends to hang to the left.
Since a single lead requires at least 2 wires to measure a polar electrical signal, this is the minimal requirement for defibrillators, and hence any electronic patient record system associated with defibrillation. For more advanced purposes... anything beyond defibrillation, more than one lead must be immediately accessible for recording purposes. Hence, any ICU/CCU/ER/Ambulance (clinical environment) will always use multiple leads, as that is the standard of care in the USA. So until this system can use multiple wireless sensors to capture multiple leads, and buffer that data for analysis and/or just raw transmission, this approach will not do much good in clinical settings in the USA. Not capturing all the raw data would also have significant issues in clinical settings.
Unfortunately, cardiac monitoring for the purposes of an electronic patient record, in clinical settings the USA, requires the raw EKG data from multiple simultaneous leads... true leads. Analysis of the data is by FDA certified software/hardware, after which data can be stored and accessed by "event or significant change from a baseline", which is also recorded on a regular basis.
The FDA hurdles for this device are significant... it would need to go through at least two rounds of FDA acceptance for two separate functions.... the physical device itself, and the software it uses for both intelligent data analysis and data transmission. These are significant hurdles. Physically, the device would need to tolerate repeated defibrillations, including direct discharge contact, and would need to function with external pacemakers. The 2.4 GHz range is very susceptible to noise, especially high powered WiFi... which begs the question of why use ANT+ in clinical settings, especially when you need a gateway to TCP/IP anyway.
Given what FDA approval would cost (millions), and given that the EKG wires are not really a hassle to clinical providers as long as they are not attached to a physical monitor (read... this means the wires are attached to a wireless hub that is attached to the patient and sends all lead data [not summaries or events] via TCP/IP [with full buffering] over WiFi, with IEEE 802.1 physical security certs and WPA2 Enterprise encryption) you will never see this device in a clinical setting in the USA as it doesn't meet clinical standards.
InterestedInRF, thanks for the clarification about where the device is sold, and especially thanks for the info about data security and the ANT protocol.
WiFi, when properly secured (e.g. with WPA2) is quite secure The reason it's inappropriate here has to do with power requirement for WiFi.
ANT is a short range protocol that is more suited to low-power app like this.
Full disclosure: I work in Nordic Semiconductor's marketing department. First, it's important to note that Lifetouch monitors are not currently being offered for commercial or clinical use in the U.S. The Isansys HRV011 is currently undergoing medical trials in a clinical environment in Europe. As in the U.S., patient confidentiality is a big deal there and transmitted data is fully encrypted. Isansys uses the ANT base protocol (see www.thisisant.com) but adds its own proprietary application layer for added security. As "Noswad" comments the only thing that's attached to the patient is the 'patch' shown in the picture at the head of the article.
I wish no one had to worry about hackers in any application, but we do. Unfortunately, data security is never a non-issue, and wireless networks are eminently hackable. At least in the US, confidentiality of patient records is a legal issue. So I'm still curious about the ANT standard and encryption.
There is a link to the manufacturer website. It is a self contained unit that transmit data to a central hub. It appears to be about 4-5" long and 1-2" wide.
What possible commercial or criminal interest could there be in hacking this data? If the patient has already been through intensive care, his/her insurance credentials are shot, so the insurance companies don't stand to gain from intercepting data. Data security is therefore a non-issue - simple packet ID would be enough.
Don't know if I missed it in the article, but how big is this? Is it just a matter of taping it on the patient and leaving it there? Does it need to be wired to sensors around the body?
By refining topologies and using new fluid technology, Moog's new peak sine drive controller increases available power without increasing controller volume.
Lantronix Inc. has expanded its line of controllers for sensor networks with the release of a rugged controller that improves management of automation systems used in a number of industries, including manufacturing, oil and gas, and chemicals.
Inspired by the hooks a parasitic worm uses to penetrate its host's intestines, the Karp Lab has invented a flexible adhesive patch covered with microneedles that adheres well to wet, soft tissues, but doesn't cause damage when removed.
From Dell / Intel® New Paradigms in Design Work Scott Hamilton, vertical market strategist for Dell Precision workstations, 5/2/2013 3
Early in my career, I worked as a draftsman and remember the days of drawing on vellum with numbered pencils and Mylar with plastic lead. This was a fun experience in the sense that I ...
I've been using workstations for more than 10 years and love finding ways to get more performance from my system. With demanding professional applications that require more power each ...
A lasting memory from my first job as an engineer in an auto assembly plant is standing on hard concrete at six in the morning, vending-machine coffee clutched in hand, listening to ...
A quick look into the merger of two powerhouse 3D printing OEMs and the new leader in rapid prototyping solutions, Stratasys. The industrial revolution is now led by 3D printing and engineers are given the opportunity to fully maximize their design capabilities, reduce their time-to-market and functionally test prototypes cheaper, faster and easier. Bruce Bradshaw, Director of Marketing in North America, will explore the large product offering and variety of materials that will help CAD designers articulate their product design with actual, physical prototypes. This broadcast will dive deep into technical information including application specific stories from real world customers and their experiences with 3D printing. 3D Printing is
To save this item to your list of favorite Design News content so you can find it later in your Profile page, click the "Save It" button next to the item.
If you found this interesting or useful, please use the links to the services below to share it with other readers. You will need a free account with each service to share an item via that service.